Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

How to invert the ppg signal #8

Open
ajmalazees opened this issue Jul 8, 2018 · 4 comments
Open

How to invert the ppg signal #8

ajmalazees opened this issue Jul 8, 2018 · 4 comments
Assignees

Comments

@ajmalazees
Copy link

The ppg signal obtained through this code is an inverted graph nah? therefore i want to invert the ppg signal t get the proper output for my project. Also my project is to calculate hemoglobin concentration, how can i modify the above code to calculate hemoglobin concentration.

I want to get the ppg signal as below

image

@adelalfusail
Copy link

Exactly I entered to ask this question the PPG wave is inverted put I thank the problem is coming from probe the red and infrared PD is not connected correctly so I am now try to open my probe and check it than I will tell you what is the result

@niv22
Copy link

niv22 commented Feb 1, 2021

I have the same concern. Is there a reason why the PPG signal is inverted?

@joicetm
Copy link

joicetm commented Feb 2, 2021

Hi niv22,

The signal from afe4490 comes by default as "inverted from other ppg publications".
It is mentioned in the datasheet on page 82. There is no option given in the afe4490 to invert the signal directly from the chip but you can use some software techniques to invert the signal if it is required.

Screenshot from 2021-02-02 10-17-51

@niv22
Copy link

niv22 commented Feb 15, 2021

@joicetm, that was helpful. Thank you.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment